Hi everyone, I have the same problem as discussed in this post 7 years ago:
The thing is, I'm student doing my final degree project with PXI 1042Q with the modules 5610, 5441, 5600, 5142 and controller 8106. As for the bad situation in Vnzla, there is no money to send PXI 5610 for repair , so I would please beg anyone for any help or procedure that could work this out.
Im getting this error in Labview from PXI 5610 at Initiate RFSG session.
DAQmx Error -200617 occurred:
Measurements: Output generation was aborted by the reverse power protection circuitry of the device. Either the output signal exceeded the output power limit, or power was being driven back into the output of the device by an external source. Correct the problem, then generate the signal again.
In MAX the self-test also fails which fails with error -200020.
Project involves transmission and reception using RFSG and RFSA. This was working perfectly days ago, suddenly today, it doesnt.